Objective. Our goal was to determine the conditions for which a more precise calculation of the electric potential than the quasi-static approximation may be needed in models of electrical neurostimulation, particularly for signals with …
        
      
     
  
    
    
    
    
      
      
        
          Objective. Models of excitable cells consider the membrane specific capacitance as a ubiquitous and constant parameter. However, experimental measurements show that the membrane capacitance declines with increasing frequency, i.e., exhibits …
        
      
     
  
    
    
    
    
      
      
        
          textcopyright 2015 IEEE. Transcutaneous electrical stimulation (TES) of nerve fibers may be optimized by overcoming the impedance of the skin using high-frequency (HF) stimulation signals. The excitation properties of nerve fibers depend on the …
        
      
     
  
    
    
    
    
      
      
        
          OBJECTIVE: Incorporating high-frequency components in transcutaneous electrical stimulation (TES) waveforms may make it possible to stimulate deeper nerve fibers since the impedance of tissue declines with increasing frequency. However, the …